- Description
- Features / Benefits
- Specifications
The VUV-24 is a high sensitivity variable wavelength light absorbance monitor designed for chromatography. It is a true dual beam instrument utilizing a deuterium light souce and prism arrangement that provides wavelengths from 190nm to 400nm.
The VUV-24 has a broad line of sample cells available. These include cells designed for on column capillary monitoring, standard and high flow 5 mm and 10 mm path length HPLC cells, and preparative LC.
VUV-24

High Sensitivity
The VUV-24 series detector has a high-resolution analog output with sensitivity range from 4.0 AU/V to 0.0005 AU/V and feature a state-of-the-art noise level of less than +/-0.00001 AU
Wavelength Selection
The deuterium lamp has a working wavelength range from 190 nm to 400 nm. Specific wavelengths are selected by using a keypad input or RS-232 interface.
Reliable, Rugged, Low Maintenance Design
The VUV-24 is constructed with a high reliability deuterium lamp, true double beam sample/reference light paths and a robust universal power supply. A rugged tactile feedback keypad for instrument control is standard.
VUV-24

| Light source | Deuterium lamp 190 nm - 400 nm, >2000 hour service life |
| Optical design | Double beam (sample/reference) |
| Wavelength selection | User selectable (front panel or RS232) in 1 nm increments |
| Wavelength accuracy | Better than +/- 2 nm range |
| Wavelength precision | Better than +/- 0.2 nm |
| Band pass | <8.0 nm (typical) at half-height @ 250 nm |
| Noise* | <+/- 1 x 10-5 AU over 30 sec period peak to peak with 1.0 sec rise time |
| Drift* | <+/- 1 x 10-4 AU over 30 min period peak to peak 1.0 sec rise time |
| Input/Output | RS-232 computer I/O; auto-injection (event or run); auto-zero input; one analog integrator or chart recorder output, (voltage range default 1.0 V/AU); selectable analog range outputs of 4.0, 2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.01, 0.005, 0.002, 0.001, 0.0005 AU/V with voltage range of +/- 4V at any scale |
| Display | Absorbance (5 digit LED); wavelength (3 digit LED); analog output range (momentarily from absorbance display by pressing range switch); reference and sample light levels |
| Size | 6.6" x 9.4" x 15" (HxWxD), Approx. 12 lb (shipping weight) |